通过硫酸亚铁电位滴定法测定AuBe1、AuNi9、AuAgCu35-5、PdAgCuAuPtZn30-14-10-10-1合金中Au的实践,分析了测定不同Au含量过程中的不确定度来源,建立了数学模型,计算了各标准不确定度分量、合成标准不确定度和扩展不确定度。结果表明:测定合金样品中9.99%、60.16%、90.85%、98.93%的Au含量,扩展不确定度分别为0.02%、0.14%、0.22%、0.24%。
The uncertainty for potentiometric titration of gold in AuBe1, AuNi9, AuAgCu35-5, PdAgCu-AuPtZn30-14-10-10-1 alloys using ferrous sulfate was evaluated. The sources of uncertainty in the process of determination of gold contents were analyzed, and the mathematical model was established also. The components of standard uncertainty, combined standard uncertainty and expanded uncertainty were calculated. Results indicated that the expanded uncertainties are 0.02%, 0.14%, 0.22% and 0.24% for gold contents of 9.99%, 60.16%, 90.85%and 98.93%in alloy samples, respectively.
参考文献
[1] | 国家质量技术监督局计量司.测量不确定度评定与表示指南[M].北京:中国计量出版社,2004 |
[2] | 国家质量技术监督局计量司.ISO/IEC 17025校准和检验实验室能力的通用要求[M].北京:中国计量出版社,2005 |
[3] | 中国实验室国家认可委员会.化学分析中不确定度的评估指南[M].北京:中国计量出版社,2002 |
[4] | 李慎安;王玉莲;范巧成.化学实验室测量不确定度[M].北京:化学工业出版社,2008 |
[5] | 陈爱平,陈志慧,王苏明,王烨.氢醌滴定法测定金矿石中金的不确定度评定[J].岩矿测试,2006(04):360-364. |
[6] | 张华,司得勇,徐忠良,伊力哈木,周新春,王开崎.活性碳富集分离-氢醌容量法测定金矿石中金的不确定度评定[J].新疆地质,2007(02):210-212. |
[7] | CSM01010101-2006.滴定法测量结果不确定度评定规范[S].北京:中国标准出版社,2006. |
[8] | 贵研铂业股份有限公司.贵金属合金化学分析方法金、铂、钯合金中金量的测定硫酸亚铁电位滴定法[M].北京:中国国家标准化管理委员会,2008 |