纵向柔顺性是支架很重要的一个力学性能,它决定了支架能否被顺利地输送到病变部位并与血管相适应.主要目的就是利用有限元法来研究支架的设计参数对柔顺性能的影响.利用多点约束单元,在支架的简化模型上均匀地加栽弯矩使支架发生弯曲变形.结果显示,S型支架的柔顺性最好,L型支架的柔顺性最差;支架支撑体宽度和波形环曲率半径对柔顺性的影响甚微;增加支架连接杆的长度能够提高支架的柔顺性,但是,增加支架连接杆的宽度或者厚度会降低支架的柔顺性.因此,有限元法对支架力学行为分析具有很大的帮助,为支架的临床选择及优化设计提供了重要指导作用.
参考文献
[1] | Julio C P .[J].Journal of ENDOVASCULAR THERAPY,2004,11:200-206. |
[2] | Helen M B;Hunter W L.[J].Advanced Drug Delivery Reviews,2006:58z 350-58z 357. |
[3] | Ormiston JA;Dixon SR;Webster MW;Ruygrok PN;Stewart JT;Minchington I;West T .Stent longitudinal flexibility: a comparison of 13 stent designs before and after balloon expansion.[J].Catheterization and cardiovascular interventions: Official journal of the Society for Cardiac Angiography & Interventions,2000(1):120-124. |
[4] | Schmidt W;Behrens P;Werner D et al.[J].Biomedizinische Technik,2005,50(z1):1505-1506. |
[5] | Schmidt W;Andresen R;Behrens P.Comparison of Mechanical Properties of Peripheral Self-expanding Nitinol and Balloon-expandable Stainless-Steel Stents[M].Barcelona,Cardiovascular and Interventional Radiological Society of Europe,2004 |
[6] | Mori K;Saito T .Effects of stent structure on stent flexibility measurements.[J].Annals of Biomedical Engineering,2005(6):733-742. |
[7] | 张庆宝,王伟强,齐民,梁栋科,杨大智.不同材料冠状动脉支架膨胀行为分析[J].功能材料,2007(01):130-134. |
[8] | Migliavacca F;Petrini L;Montanari V et al.[J].Medical Engineering and Physics,2005,27:13-18. |
[9] | Wu W;Qi M;Liu XP;Yang DZ;Wang WQ .Delivery and release of nitinol stent in carotid artery and their interactions: a finite element analysis.[J].Journal of Biomechanics,2007(13):3034-3040. |
[10] | Mortier P;Beule M D;Carlier S G et al.[J].Journal of Biomechanical Engineering,2008,130:1-7. |
[11] | Petrini L.;Migliavacca F.;Auricchio F.;Dubini G. .Numerical investigation of the intravascular coronary stent flexibility[J].Journal of Biomechanics,2004(4):495-501. |
[12] | Etare F;Finer G;Boivin M et al.[J].Journal of Biomechanics,2001,34:1065-1075. |
[13] | 高振宇,梁栋科,齐民,张庆宝,杨大智.镍钛合金超弹性支架纵向柔顺性分析[J].功能材料,2007(01):113-114. |
[14] | Wu W;Yang D Z;Wang W Q .[J].Journal of Materials Processing Technology,2007,184:447-450. |
[15] | Takashima K;Kitou T;Mori K;Ikeuchi K .Simulation and experimental observation of contact conditions between stents and artery models.[J].Medical engineering and physics,2007(3):326-335. |