以四乙氧基硅烷(TEOS)和二甲基二乙氧基硅烷(DDS)为共聚前驱体,Ru(phen)3Cl2为荧光指示剂,采用溶胶凝胶技术,通过改变DDS/TEOS比值,制备了基于荧光猝灭原理的有机改性氧敏感膜.探讨了成膜条件,研究了DDS的掺杂量对敏感膜结构和性能的影响.结果表明,掺入适量的有机硅醇盐DDS,敏感膜均匀无龟裂,有较好的机械性和柔韧性.将敏感膜组装到基于相移检测的光纤氧传感器并对气态氧进行检测,检测下限达2×10-6,响应时间<30 s;对溶解氧进行检测,检测下限达0.5 mg/L,响应时间<50 s.与纯TEOS敏感膜相比,有机改性敏感膜具有更好的重复性和稳定性,更有利于溶解氧的测量.
参考文献
[1] | MacCraith B D;O'Keeffe G;McEvoy A K .[J].Sensors and Actuators B-chemical,1995,29:51. |
[2] | 李清文,王义明,罗国安.溶胶凝胶技术在生物传感器中的应用[J].化学通报,2000(05):14-19. |
[3] | 姜德生,刘尔,陈兴,黄俊.Design and properties study of fiber optic glucose biosensor[J].中国光学快报(英文版),2003(02):108-110. |
[4] | Zhu Y X;Ran Y;Xu S Q .[J].Spie,2001,4414:137. |
[5] | Costa-Fernandez JM.;Sanz-Medel A.;Diaz-Garcia ME. .Sol-gel immobilized room-temperature phosphorescent metal-chelate as luminescent oxygen sensing material[J].Analytica chimica acta,1998(1/3):17-26. |
[6] | Remillard J T;Jones J R;Poindexter B D et al.[J].Applied Optics,1999,38:5306. |
[7] | Choi MMF.;Xiao D. .Oxygen-sensitive reverse-phase optode membrane using silica gel-adsorbed ruthenium(II) complex embedded in gelatin film[J].Analytica chimica acta,1999(2):197-205. |
[8] | Huang J;Han Y;Yue F Y et al.[J].Key Engineering Materials,2003,249:421. |
[9] | Mcevoy AK.;Maccraith BD.;Mcdonagh CM. .DISSOLVED OXYGEN SENSOR BASED ON FLUORESCENCE QUENCHING OF OXYGEN-SENSITIVE RUTHENIUM COMPLEXES IMMOBILIZED IN SOL-GEL-DERIVED POROUS SILICA COATINGS[J].The Analyst: The Analytical Journal of the Royal Society of Chemistry: A Monthly International Publication Dealing with All Branches of Analytical Chemistry,1996(6):785-788. |
[10] | McDonagh C M;MacCraith B D;McEvoy A K .[J].Analytical Chemistry,1998,70:45. |
[11] | Sprintschnik G;Sprintschnik H W;Kirsch P P et al.[J].Journal of the American Chemical Society,1977,99:4947. |
[12] | 中国医学科学院卫生研究所.水质分析法[M].北京:人民卫生出版社,1983 |
[13] | Jiang D S;Chen X;Liu E et al.[J].J Wuhan Univ Tech-Mater Sci,2002,17(02):51. |
[14] | 曹冰;朱从善 .[J].中国科学E辑,1998,28:394. |