A differential pulse voltammetric method was developed for the sensitive and selective determination of Cu(Ⅱ) at 4-methoxy-2,6-bis(3,5-dimethylpyrazoyl)-1,3,5-triazine modified carbon paste electrode in 0.05 mol/L KHC8H4O4 solution (pH = 4.02). The oxidation peak of Cu(Ⅱ) was observed at 0.065 V(vs Ag/AgC1) by scanning the potentid in positive direction. The analysis procedure consisted of an open circuit accumulation step in stirred sample solution. It was followed by medium exchange to a clean solution and subsequently an anodic potential scan was affected to obtain the voltammetric peak. The current was proportional to the concentration of the Cu(Ⅱ) ion in a range of 1 × 10-7 -1 × 10 -4 mol/L for 6 min accumulation; the most of metal ions did not interfere with the determination. The developed method was applied to Cu(Ⅱ)determination in coal-ash sample, the results agreed with that of atomic adsorption spectroscopy(AAS).
参考文献
[1] | Wang J.Analytical Electrochemistry[M].VCH Publishers,New York,1994 |
[2] | Kalcher K;Grabec I;Raber G;Cai X., Tavcar G .and Ogorevc B.,The vermiculitemodified carbon paste electrode as a model system for preconcentrating mono- and divalent cations[J].Electroanalytical chemistry,1995,386:149. |
[3] | Zhang H L;Yie Y·K;Xu B .Preparing of α-benzil dioxmie modified carbon paste electrode and determination of copper in water samples [J].Analysis Laboratory(in Chinese),1999,18(05):80. |
[4] | Lan W H;Lu G H;Yao S L .Determination of trace copper by chitin modified carbon paste electrode[J].Analysis Chemistry(in chinese),1998,26(10):1192. |
[5] | Xibin Zhou.The Application of Modified Electrode by Nitrc-heterocyclic combination[DJ[J].Northwest Normal University,LanZhou,1999:30. |