欢迎登录材料期刊网

材料期刊网

高级检索

建立了王水溶解纯金悪品、标准基体匹配、电感耦合等离子体发射光谱(ICP-AES)测定纯金饰品中痕量钯元素的定量分析方法。该方法具有简便、快速、灵敏度高、重现性好等优点。方法检出限为0.08μg/g,加标回收率为104%,相对标准偏差(RSD)为7.52%,A类不确定度为0.0119 mg/L。

The quantitative analysis method of trace palladium element in pure golden jewelry with inductively coupled plasma- atomic emission spectrometry (ICP-AES) was established. The samples were dissolved in aqua regia and Pd solution was matched with gold matrix. And then detected by ICP-AES. The test method proved better sensitivity,lower uncertainty and good accuracy, but also the test speed can be improved enormously. The analytical results showed that the detection limit of palladium trace element was 0.08 μg/g.The recovery rates of the studied elements was 104%.The relative standard deviations (RSD) of the method was 7.52%. The uncertainties of type A was 0.0119 mg/L.

参考文献

[1] GB/T 11066.6-2009.金化学分析方法镁、镍、锰和钯量的测定火焰原子吸收光谱法[S].北京:中国标准出版社,2009.
[2] GB/T11066.8-2009.金化学分析方法银、铜、铁、铅、锑、铋、钯、镁、镍、锰和铬量的测定乙酸乙酯萃取-电感耦合等离子体原子发射光谱法[S].北京:中国标准出版社,2009.
[3] 刘荣,梁杰,夏成强.ICP-AES法直接测定纯金中的杂质元素[J].黄金,2008(07):54-55.
[4] 陆文伟,邢培儒.电感耦合等离子体发射光谱分析金中杂质元素[J].贵金属,1997(02):37-41.
[5] 田治龙,李中宇.用ICP-AES分析高纯黄金中的杂质[J].黄金,2004(11):53-56.
[6] 臧慕文.ICP-AES法测定钯碳催化剂中钯的不确定度的评定[J].分析试验室,2003(06):13-17.
[7] 曹宏燕.分析测试中测量不确定度及评定第五部分测量不确定度评定中要注意的一些问题[J].冶金分析,2006(04):89-94.
[8] 王铮,许实,邹亚娟.ICP-AES测试玩具样品中铅的不确定度研究[J].光谱实验室,2008(05):905-908.
[9] 朱晨燕.ICP-OES测定水中锰的结果不确定度评定[J].中国给水排水,2008(04):86-88.
[10] 张凤霞,李桂华,程佑法,山广祺,李婷.ICP-AES法测定千足黄金饰品中银含量的不确定度评估[J].贵金属,2012(02):63-66.
[11] JJF 1059-1999.测量不确定度评定与表示[S].北京:中国计量出版社,1999.
[12] JJF 1135-2005.化学分析测量不确定度评定[S].北京:中国计量出版社,2005.
上一张 下一张
上一张 下一张
计量
  • 下载量()
  • 访问量()
文章评分
  • 您的评分:
  • 1
    0%
  • 2
    0%
  • 3
    0%
  • 4
    0%
  • 5
    0%