系统地研究了影响冠脉支架激光精细雕刻切口质量的因素,包括激光器切割工艺参数:波长1064nm,光斑直径2~3mm,脉宽12~15μs,脉冲频率1000~1200Hz,平均功率15W,脉冲能量10~15mJ;切割速度:30cm/min;辅助气体O2压力0.2~0.4MPa;喷嘴与工件表面距离L=4.09mm.建立了激光切割过程温度场的数学模型.结果表明冠脉支架筋宽理论值与实际切割值存在±0.01mm误差,由阿贝-赫梅特判据判定激光切割系统存在周期性系统误差.确定了316L不锈钢冠脉支架Nd-YAG激光器精细加工工艺参数,实现了冠状动脉支架精度为0.01mm的激光精密切割.

参考文献

[1] 张岳,郝丽,柳华,刘伟奇.激光显示的原理与实现[J].光学 精密工程,2006(03):402-405.
[2] 杨超君,周建忠,张永康,张昉.激光冲击金属板料变形的最小激光能量估算及其实验研究[J].光学 精密工程,2006(03):396-401.
[3] 李凤友,李红军,卢振武,廖江红.激光直写系统焦斑整形的研究[J].光学精密工程,2001(01):14-18.
[4] 卢启鹏,侯宝忠,朱琳琳.半导体激光血管内照射治疗静脉曲张的机理研究[J].光学精密工程,2006(04):589-593.
[5] 毛英泰.误差理论与精度分析[M].北京:国防工业出版社,1982
[6] Dotter C T;Judkins M P .Transluminal treatment of arteriosclerotic obstruction.Description of a new technique and a preliminary report of its application[J].Circulation,1964,30:654.
[7] Dotter C T;Buschmann R W;McKiuney M K et al.Transluminal expandable nitinol coil stent grafting:preliminary report[J].Radiology,1983,147:259.
[8] Palmaz J C et al.Expandable intraluminal graft:a preliminary study[J].Radiology,1985,156:63.
[9] Pukin L .Devices and techniques for endovascular surgery: catheters, stents, coated stents, and stented grafts.[J].The Mount Sinai journal of medicine,2003(6):386-392.
[10] Schatz R A;Palmaz J C;Tio F O et al.Balloon expandable intracoronary stents in the adult dog[J].Circulation,1987,76:450.
[11] Baleon R .Recommendations on steut manufacture,implantation and utilization[J].European Heart Journal,1997,18:1536.
[12] Serruys P W;de Jaegere P;Kiemene0 F et al.The BENESTENT Study Group.A comparison of balloonexpandable-stent implantation with balloon angioplasty in patients with coronary artery disease[J].New England Journal of Medicine,1994,331:489.
[13] Fischman D L;Leon M B;Baim D S et al.The Stent Restenosis Study Investigators.A randomized comparison of coronary stent placement and balloon angioplasty in the treatment of coronary artery disease[J].New England Journal of Medicine,1994,331:496.
[14] Colombo A;Hall P;Nakamura S et al.Intracoronary stenting without anticoagulation accomplished with intravascular ultrasound guidance[J].Circulation,1995,91:1676.
[15] Duerig T W .An overview of superelastic stent design[J].Minimally Invasive Therapy and Allied Technologies,2000,9(34):235.
上一张 下一张
上一张 下一张
计量
  • 下载量()
  • 访问量()
文章评分
  • 您的评分:
  • 1
    0%
  • 2
    0%
  • 3
    0%
  • 4
    0%
  • 5
    0%