采用微芯片毛细管电泳非接触电导检测法,对黄花鱼中非法添加的工业染料碱性嫩黄O进行了分析.探讨了缓冲液种类、浓度,分离电压和进样时间等因素对分离检测的影响.实验选择5.0 mmol/L乳酸缓冲液(pH =3.29)、1.8 kV分离电压,在1.0 min内实现了碱性嫩黄O的快速分离测定.在优化条件下,碱性嫩黄O浓度的线性范围为5.0 ~ 100.0 mg/L,黄花鱼中碱性嫩黄O的检出限为0.2 mg/kg,该法可成功测定黄花鱼中碱性嫩黄O的含量.
参考文献
[1] | LU Shiying,ZOU Mingqiang.The Hazards of Frequent Inedible Colorants Used in Foods and Their Detection Methods[J].China Instrum,2009,8:5-50 (in Chinese).卢士英,邹明强.食品中常见的非食用色素的危害与检测.[J].中国仪器仪表,2009,8:5-50. |
[2] | NIE Xuemei,CHU Xiaogang,SUN Huijie,et al.Study on the Use of Dyes in EU food[J].Sci Technol Food Ind,2010,31(4):392-398 (in Chinese).聂雪梅,储晓刚,孙惠杰,等.欧盟食品中染料使用情况的研究[J].食品工业科技,2010,31(4):392-398. |
[3] | LIN Saijun,TU Haiyun,SUN Lan,et al.Simultaneous Determination of Five Yellow Dyes in Foods by High Performance Liquid Chromatography Coupled with Tandem Mass Spectrometry[J].Chinese J Chromatogr,2011,29 (1):79-82 (in Chinese)林赛君,屠海云,孙岚,等.高效液相色谱-串联质谱法同时测定食品中五种黄色化工染料[J].色谱,2011,29(1):79-82. |
[4] | LIN Qin.Simultaneous Determination of Auramine O and Chrysoidine in Bean Products by HPLC[J].Chinese J Chromatogr,2007,25 (5):776-777 (in Chinese).林钦.高效液相色谱法同时测定豆制品中的碱性橙和碱性嫩黄O染料[J].色谱,2007,25 (5):776-777. |
[5] | ZHANG Haiqi,LIANG Lijun,HE Zhongyang,et al.Determination of Auramine O in Fishery Products by High-Performance Liquid Chromatography-Electrospray Tandem Mass Spectrometry[J].J Chinese Mass Spectrosc Soc,2010,31(1):49-52(in Chinese).张海琪,梁丽军,何中央,等.水产品中碱性嫩黄O残留量的液相色谱-串联质谱测定[J].质谱学报,2010,31(1):49-52. |
[6] | ZHENG Xiaoyan.Simultaneous Determination of Chrysoidine,Auramine O and Safranine T in Foods by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry[J].J Anal Sci,2009,25 (4):409-415 (in Chinese).郑小严.超高效液相色谱-串联质谱法同时测定食品中碱性橙、碱性嫩黄O和碱性桃红T[J].分析科学学报,2009,25(4):409-415. |
[7] | YANG Xiujuan,LI Oulian,CHEN Zuanguang,et al.Determination of Pseudoephedrine Hydrochloride and Dextromethorphan Hydrobromide in Cold Tablet by Micro-fluidic Chip[J].Chinese J Anal Chem,2008,36 (5):673-677 (in Chinese).杨秀娟,李偶连,陈缵光,等.微流控芯片非接触电导法测定感冒药中的盐酸伪麻黄碱和氢溴酸右美沙芬[J].分析化学,2008,36(5):673-677. |
[8] | Chen Z G,Li Q W,Li O L,et al.A Thin Cover Glass Chip for Contactless Conductivity Detection in Microchip Capillary Electrophoresis[J].Talanta,2007,71 (5):1944-2195. |
上一张
下一张
上一张
下一张
计量
- 下载量()
- 访问量()
文章评分
- 您的评分:
-
10%
-
20%
-
30%
-
40%
-
50%