目前地质样品中的痕量银一般采用双电极发射光谱法进行测定,但方法流程长、人力物力消耗大、数据稳定性差,且很难满足银检出限的要求。为了解决这些问题,该文以硝酸-高氯酸-氢氟酸溶解试样,2%的盐酸浸取,2%的磷酸氢二铵-2%硫脲溶液作为基体改进剂,以石墨炉原子吸收法测定地质样品中痕量银;采用该方法对国家一级标准物质及实际样品进行了分析,获得了令人满意的结果;方法的检出限为0.02μg/g,加入标准物质回收率为95.0%~98.3%,相对标准偏差为4.56%~9.98%。该试液体系可应用于其他元素测定,无需再溶解样品,简化了分析流程,降低了测试成本。该方法简便、快速、准确。

Dual electrodes emission spectroscopy commonly used in the determination of trace silver in geological samples is lengthy in flowsheet and consumes much man power and lots of material resources besides its poor data con-sistency and detection limit.To solve these problems,the research dissolves samples using nitric acid-perchloric acid-hydrofluoric acid,uses 2 % hydrochloric acid for leaching,2 % diammonium hydrogen phosphate -2 % thiorea as matrix modifier,and finally uses graphite furnace atomic absorption to determine trace silver in geological samples;the method analyzed national primary reference substance and field samples and the results are satisfactory;the detection limit is 0.02 μg/g,the recovery rate adding reference substance is 95.0 % to 98.3 % and the RSD is 4.56 % to 9.98 %.The test solvent regime is applicable to the determination of other elements with no need to additionally dis-solve samples which makes analysis simple and cost-effective.The method is simple,quick and accurate.

参考文献

[1] 张志龙.有色地质分析规程[M].北京:中国有色金属工业总公司地质局,1992:41-49.,1992.
[2] 叶家瑜,江宝林.区域地球化学勘查样品分析方法[M].北京:地质出版社,2004:123,352.,2004.
[3] 刘红毅,罗大芳,鲍爱华.GFAAS法测定地质样品中痕量银[J].云南地质,2011(03):348-352.
[4] 薛光.银的分析化学[M].北京:科学出版社,1988:382-383.,1988.
[5] 李建 .钴和硝酸镁作为通用基体改进剂的研究[J].理化检测:化学分册,1997,33(4):160-163.,1997.
[6] 李维,余林成.石墨炉原子化光谱法连续测定饮水中6种金属含量[J].中国卫生检验杂志,2011(08):2094-2095.
[7] 牛占海,林焰,和督虎.石墨炉原子吸收法测定化探样品中银和镉[J].冶金分析,2001(06):61-63.
[8] 王新华.以吡咯啶二硫代氨基甲酸铵为基体改进剂的水中银的石墨炉原子吸收测定法[J].环境与健康杂志,2010(10):895-897.